24.8.2. Time interval between changing dosage thresholds
By setting <P7.4.2 DELAY> parameter users can set a delay time
between changing dosing thresholds.
Procedure:
•
Enter
<P7 WORK MODES> and then:
•
Using numeric keys inscribe the required value (0 to 60 s)
and confirm by pressing
.
Return to weighing:
See 11.3.
24.8.3. Time interval completing process
Users can set <P7.4.3 CHUTE TIME> parameter to change time
of completing a process of dosing after indication drops below LO.
Procedure:
•
Enter
<P7 WORK MODES> and then: