- 96 -
17.2.2.
Setting the reference mass by determining from the sample
of known quantity
While determining the mass of a single piece the function
ACAI
(Automatic
Accuracy Correction) is in use.
Means of operation of ACAI function:
-
Number of pieces ( on adding) on balance’s weighing pan has to be
greater than has been previously
-
Number of pieces (on adding) on balance’s weighing pan must be less
than twice the amount of which was visible on the display before
adding parts
-
Current quantity of parts must be within the
±
0,3 tolerance of the total
value,
-
Measurement result has to be stabilised.
Procedure:
•
Place the container on the pan and tare its mass,
•
Press one of the F buttons, to which the function <
ENTER SAMPLE
>
is assigned, and the editing window <
SAMPLE QUANTITY
> is
displayed
•
Use the navigating buttons
or
to select the correct
sample quantity.
•
For the option: any quantity (displayed value <0000> enter a numer
using navigating buttons).