- 159 -
Send stable measurement result in basic measuring unit
Format:
S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement
result
S_I CR LF
- command understood but not accessible at this moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Frame format:
1
2-3
4
5
6
7-15
16
17 18 19 20 21
S space stability
marker space character mass space
unit
CR LF
Example:
S CR LF
– command sent from a computer
S _ A CR LF
– command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
– command carried out,
response in mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format:
SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response is immediate with mass value in basic weighing
unit
Frame format:
1 2
3
4
5
6
7-15
16
17 18 19 20 21
S I space stability
marker space character mass space
unit
CR LF
Example:
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
– command carried out,
immediate response of mass value in basic weighing unit.