Start the proces.
•
Following the displayed messages set the correct options.
•
Enter the ambient temperature, next press ENTER for confirmation.
The computer software proceeds to the next step, i.e. ambient
humidity settings.
•
Set the ambient humidity, next press ENTER for confirmation. The
computer software proceeds to the next step, i.e. air pressure
settings.
•
Set the air pressure, next press ENTER for confirmation. The
computer software proceeds to the next step, i.e. entering the first
control volume (V1) for a tested pipette.