- 62 -
•
P1.4
CALIB. TEST
– Adjustment test
This fuction enables comparing the result of internal automatic adjustment
with the value of internal weight saved in balance’s factory parameters.
The process is conducted automatically and its result is shown on the
display (if the balance is connected via RS 232 port with a computer or
printer, the adjustement result will be printed out). Press
ESC
to go back to
the previous menu.
•
P1.5
AUTO. CALIB. T.
– Temperature of automatic adjustment
(balances with internal adjustment)
Determination of factor which signals start of automatic internal adjustment
NONE
-
no
factor
will
cause
start
of
adjustement
(adjustment automatically turned off)
TEMP.
- adjustment triggered by change of temperature
TIME
- adjustment triggered by time interval set in
P1.6 AUTO CALIB.
C.
BOTH
- adjustment triggered by time and temperature.
CAUTION
Function unavailable in non-verified balances.
•
P1.6
AUTO CALIB. C. - Time of automatic adjustment (balances
with internal adjustment system)
Determination of time interval, after which automatic adjustment process is
initiated.
CAUTION
Function unavailable in verified balances.