- 105 -
•
Use the arrow buttons to set the tolerance value
•
Confirm the entered tolerance by pressing
button,
•
The balance software automatically enters the mode <
DOSING
> and
displays the target mass value with a minus sign, and in the bottom
line standard mass value – target mass (if such option for the function
<
INFORMATION
> was selected).
•
The sign <Min> is shown at the top of the display. It indicates the status of
the mass on the pan with relations to the target mass, below the value
<TARGET MASS – TOLERANCE>
<Min>: mass lower than the
Target Value - Tolerance
<Ok>: mass contained within
tolerance
Target Value +/- Tolerance
<Max>: mass higher than the
Target Value + Tolerance