- 41 -
10.4.1. Setting level of filtering
Depending on the balance working conditions you should set the filter. In ideal
conditions you can set the filter to be very fast (parameter value P.2.2.1.1 Filter to
be V.FAST); however, if the working conditions are bad (shakes, drafts), the filter
should be set to slow or very slow (parameter value P.2.2.1.1 to SLOW or V.
SLOW). The effectiveness of the filter is different for the weighing range. The filter
works slower when “approaching” the weighed mass, it works more intensively
when the weighed mass is placed within the set range of the filter (parameter filter
range is accessible only from the service menu – the user does not have an access
to it).
Depending on the adjusted filter, the weighing time will be shorter (V.FAST and
FAST) or longer (SLOW and V. SLOW)
CAUTION
The higher level of filtering, the longer the weighing time.
10.4.2. Value release
Due to various ambient conditions at workplaces, and in order to set balance to
present conditions, it is necessary to determine the value release parameter
as:
FAST.+REL., FAST
or
RELIABLE
. Depending on selected option,
weighing time is either shorter or longer.