- 147 -
19.2. Printer
In the submenu <PRINTER> the balance’s user has a possibility of choosing a port
and device, to which the data is sent by pressing
button on balance. The
content of the data being transferred is set in parameter <PRINTOUTS/PRINTOUT
GLP>.
Procedure:
−
press
button
−
enter the menu <P4 DEVICES>
−
and then enter in menu group<P4.2 PRINTER>
−
set the balance parameters related to the cooperation with the
printer
<P4.2.1 PORT> to which the printer is to be connected
Accessible options:
COM 1
or
COM 2
– RS 232 port, to which the printer is connected
USB
type A – USB port, to which PCL printer or EPSON printer is
connected
WIFI
–WIFI port, which can send data to special software
manufactured by RADWAG e.g. WIN measurement, opened on the
computer connected to the balance via WIFI.
USB flash drive
–USB port type A, to which external memory flash
drive is connected
USB PC
–USB port type B, to which the computer with the special
software by RADWAG is connected e.g. WIN measurement
Example of the measurement printout is described in the PRINT-OUTS section.
Additionaly the user can send a controlling code (of a hexadecimal form) to a
printer either at the beginning of the printout - <P4.2.2 PREFIX> or at the end of it -
<P4.2.3 SUFFIX>. Sending these codes allows to control globally both, information
and actions carried out at the beginning and/or at the end of each printout sent from
the balance to the printer.
This function will be most frequently used to send an information about the code
page of a printout sent by a balance, at the beginning, and to send a commamd
enabling the crop of a paper in EPSON printers (if the printer is equipped witht an
autocutter blade), at the end.
<PREFIX> and <SUFFIX> parameters settings are available for all the printouts
sent from balance, e.g. calibration reports, density, statistics etc. and for the
header, footer and GLP printouts.
CAUTION:
It must be remembered that inserting paper crop command to <SUFFIX>
parameter (control code) will result in sending the code after each printout. If the