AED pads are single-use only
Do not reuse AED pads. Also always keep the AED package
sealed prior to the actual use. If the sealed package is
damaged or opened, exchange it to new one immediately
Interference may be caused by implanted pacemaker
With some pacemakers, the AED may not advise an electric
shock
•
Do not attach the AED pads directly on an implanted
device
•
Attach the AED pads at least 5cm away from any
implanted device
Temperature requirement for stable operation/storage
Battery and AED pads of HR-501 can be stored from -
0
℃
to 40
℃
for short term period(under 5 days), but
recommends at least 1 hour of storage at 20
℃
~ 30
℃
temperature condition prior to actual AED operation.
For long-term storage(over 5 days), RADIAN recommends
maintaining AED with temperature range from 20
℃
to
30
℃
. AED pads and battery may have significant
performance degrade with the long-term storage(over 5 days)
at out of range of 0
℃
~40
℃
temperature condition.
Do not expose AED to water or liquid directly
IP54 grade is not capable of blocking direct inflow of liquid
completely. In case of water or liquid inflow inside the AED,
there may cause malfunction or failure of the AED
Operator is to be well-informed of method of AED operation
and CPR in advance
Some national standards restrict AED to be sold by or on the
order of a physician or practitioner licensed by law in which
operator practices to use or order the use of the AED