Temperature Range: 0 ~ 40
℃
(32 ~ 104F)
Humidity Range: 5% ~ 95% (No saturated place)
④
Avoid AED storage in below places
Place where wide temperature changes happen frequently
Place where is exposed to direct sunlight
Wet or flood places
Places close to heater or heavy electrical devices
Places where excessive shock or shake may occur
Places close to flammable materials
Places exposed to the extreme environmental weather condition
5.1.6
Battery and AED pads maintenance
①
Battery maintenance
If the battery is low, AED may not be operated properly in emergency
situation. When status LED blinks ‘
RED
’, battery pack shall be replaced
immediately. Periodically
check manufacturing date indicated on battery
label, and replace to new one if it is expired (4 years from manufacturing
date)
Replacing battery pack
Pull out the battery pack to be replaced on backside of the AED,
by pressing dual lever switch on each side of the battery pack
Push in the new battery pack facing the label below until ‘clack’
sound emits