Electric shock is not needed
①
Audio guidance: ‘No shock is needed”
②
Audio guidance: ‘If there’s no pulse, please start CPR
immediately’
During AED conducts CPR guide, no ECG analysis is proceeded.
Once CPR guidance completed, the product informs the rescue
progress status via audio message and re-analyzes ECG
Other conditions
In case of AED pads’ mis-attachment to the patient, the product
operates as below
①
Audio guidance: ‘Firmly attach the pads to the patient’s bare
skin’
Even though electric energy charge is completed, if user does
not push blinking shock button for more than 15 seconds,
AED will conduct discharge the electric energy instead of
delivering electric shock to patient through AED pads
②
Audio guidance: ‘Shock button was not pressed’
Be careful of electric shock
Electric shock may be delivered to a person who is physically
contacting with the patient. Before pushing the shock button,
confirm and instruct others to stay away from the patient