7.5
Parameters for P-303 (AED Pads)
Parameter
Detail
Type
Pre-gelled, self-adhesive, disposable, non-
polarized defibrillation pads
Shelf life
2 years
Disposal
Abide by regional/national regulations
7.6
Parameters for BT-303 (Battery)
Parameter
Detail
Type
Lithium Manganese Dioxide (LiMnO
2
)
Output Voltage
12 VDC 4.2Ah
Disposal
Abide by regional/national regulations
Shelf life
4 years from date of manufacture
(can be reduced by temperature condition)
7.7
FFT ECG Analysis
FFT ECG Analysis algorithm is for efficient and accurate ECG detection algorithm
that meets below requirements and recommendations
•
Shockable Rhythm-VF: IEC 60601-2-4, AHA’s recommend: >90%
•
Shockable Rhythm-VT: IEC 60601-2-4, AHA’s recommend: >75%
•
Non-shockable—Rhythm-NSR: IEC 60601-2-4, AHA’s recommend: >95%
•
Non-shockable-Asystole: IEC 60601-2-4, AHA’s recommend: >95%
•
Non-shockable-all other rhythms: IEC 60601-2-4, AHA’s recommend: >95%