5.2
After first-time use
If AED were operated for an actual emergency situation, AED maintenance manager
shall check-up followings to make AED ready to be used again anytime for everyone.
5.2.1
Exterior inspection after AED operation
Check exterior appearance of AED, and if it is necessary, please remove
contamination. For damage recovery on AED, please contact customer
technical support
5.2.2
Components management after AED operation
If AED pads were used or unpacked, please replace the pads to new ones
If AED delivered electric shocks at least over 3 times, and life time is
expired (2 years from manufacturing date), RADIAN suggests replacing
the battery to new one. Suggested battery life is 4 years for standby mode
6
Troubleshooting
6.1
Self-diagnosis
Power-On Self-test
(Operation mode)
Every time when power button is pressed
Flash ROM test
Charge/Discharge test
Battery Output level test
Regular Self-test
(Stand-by mode)
Automatically performed once per month on 1
st
day
Flash ROM test
Charge/Discharge test
Battery Output level test
If the product fails in self-diagnosis test, status light blinks in red and beeper sounds a
warning. In this state, the product does not operate properly. If the product does not
operate properly, users should refer to 6.3. Troubleshooting in the manual.
During periodic self-diagnosis, if product trouble is found, status light blinks in red;
in case of power on by pushing power button, beeper sounds 3 times every second for
warning with error audio guidance