Do not attempt to recharge or cause heat/damage the battery
(BT-303)
To avoid possible fire or explosion hazard, do not use around
the following situation/place.
In the presence of flammable gases
In the presence of hyperbaric oxygen tank (within 10m)
In the presence of hydrogen tank (within 10m)
Do not use battery not provided by RADIAN
RADIAN is not responsible for damage/malfunction in
product operation caused by use of unofficial battery. Since
use of unofficial battery may cause trouble/malfunction
②
AED pads maintenance
AED pads’ expiration date shall be checked periodically. Expired
AED pads shall be replaced to new one immediately
If sealing package of AED pads is ripped or damaged, do not use the
pads and replace it to new one
HR-501 only accepts P-303 AED pads, do not use any other AED
pads not provided by RADIAN
RADIAN recommends preparing additional pads with different
expiration dates for spare