4.3.3
Data management
4.3.3.1
ECG data record
The product automatically saves ECG data based on every events
of the AED operates. Those data can be transferred via Bluetooth
2.0 interface to a computer. Pairing may be required to connect
AED to computer
4.3.3.2
Heart Guardian
TM
Heart Guardian
TM
is a viewing software provided by RADIAN for
displaying and analyzing the ECG data transferred to computer.
Heart Guardian software will be updated periodically and uploaded
to RADIAN’s website, www.radian.co.kr
5
Guidance for maintenance
5.1
AED location and periodical maintenance
5.1.1
Please set AED at a place where AED can be easily detected and accessed.
RADIAN strongly suggests setting AED close to a telephone in order to call
emergency aid service immediately in case of emergency
5.1.2
All components including P-303 electric pads shall be kept together
5.1.3
Always, battery shall be installed to AED
5.1.4
Periodically check AED operation and expiration date of battery and AED pads
5.1.5
The product shall be stored/operated in following environmental conditions
①
Long-term storage condition (over 5 days)
Temperature Range: 20 ~ 30
℃
(68 ~ 86
℉
)
Humidity Range: 5% ~ 95% (No saturated place)
②
Short-term storage condition (under 5 days)
Temperature Range: 0 ~ 40
℃
(32 ~ 104
℉
)
Prior to use AED, AED shall be kept in a place at 20 ~ 30
℃
(68 ~
86
℉
) for at least 1 hour
Humidity Range: 5% ~ 95% (No saturated place)
③
Environmental requirement for AED operation