QUEST SPECTRUM
®
3
p. 45/50
Instructions for Use
RF
Common
mode/
Conducted
Susceptibil
ity
IEC
61000-4-6
Radiated
RF
Electroma
gnetic
Field
IEC
61000-4-3
3 Vrms
150 kHz to 80 MHz
6 Vrms
Inside IMS bands
3 V/m
80 MHz to 2.7 GHz
3 Vrms
150 kHz to 80 MHz
6 Vrms
Inside IMS bands
3 V/m
80 MHz to 2.7 GHz
Portable and mobile RF communications equipment should
be used no closer to any part of the QUEST SPECTRUM
®
3,
including cables than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
=1,2 √P
d
=1,2 √P 80 MHz to 800 MHz
d
=2,4 √P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Proximity
fields form
RF
wireless
communic
ations
equipment
385 MHz
380
– 390 MHz
430
– 470 MHz
28 V/m
704
– 787 MHz
9 V/m
800
– 960 MHz
28 V/m
1700
– 1990 MHz
28 V/m
2400
– 2570 MHz
28 V/m
5100
– 5800 MHz
9 V/m
385 MHz
380
– 390 MHz
430
– 470 MHz
28 V/m
704
– 787 MHz
9 V/m
800
– 960 MHz
28 V/m
1700
– 1990 MHz
28 V/m
2400
– 2570 MHz
28 V/m
5100
– 5800 MHz
9 V/m
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the QUEST SPECTRUM
®
3 is used exceeds the applicable RF compliance level above, the QUEST SPECTRUM
®
3 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the QUEST SPECTRUM
®
3.
Содержание SPECTRUM 3
Страница 1: ...Instructions for Use QUEST SPECTRUM 3 ...
Страница 19: ...QUEST SPECTRUM 3 p 16 50 Instructions for Use QUEST SPECTRUM IPC Trolley Barco Branded 4K UHD surgical display ...
Страница 20: ...QUEST SPECTRUM 3 p 17 50 Instructions for Use Instructions for Use Draping Instructions Quick Start ...
Страница 26: ...QUEST SPECTRUM 3 p 23 50 Instructions for Use Figure 6 home screen Figure 7 insert patient data ...
Страница 27: ...QUEST SPECTRUM 3 p 24 50 Instructions for Use Figure 8 Start Surgery Figure 9 Select surgery ...