QUEST SPECTRUM
®
3
p. 25/50
Instructions for Use
Focus
The camera can be focused manually by using the following options:
•
By using the
camera (see chapter 8.2 “using the camera”)
•
By using the touchpad of the IPC
When focusing, the icon will be set to MF (Manual Focus), when done this
will be set to the other icon.
Using the touchpad to focus
Select the icon on the touchpad of the IPC (see the first image on the left).
When selected, the icon will transform into an icon as shown on the left.
Focusing can be done by selecting the arrows on the side. When done
focusing, select the icon again and it will transform back into his previous
form.
Mode
When starting surgery, the correct mode needs to be selected otherwise
no fluorescent overlay can be seen. There are two modes:
•
Green icon: the 800 nm channel (supports the use of ICG)
•
Blue icon: the 700 nm channel
It is important to know what kind of fluorescence is used before starting
surgery in order for the correct mode to be selected.
Note:
Only the ICG mode is activated for USA
purposes.
Sensitivity
The sensitivity can be increased by use of this function. When unable to
clearly visualise the fluorescence during the procedure, this setting can
be
adjusted
to
improve
the
device’s
sensitivity.
The sensitivity has three different modes, the arrow represents the
sensitivity of the device within tissue, by increasing the sensitivity the
arrow “penetrates” deeper through the tissue to symbolise the increase of
the sensitivity
Note:
Consequently, when increasing the
sensitivity,
the
real-time
behavior
decreases
and
the
image
noise
increases. This may possibly result in a
lower quality image.
Содержание SPECTRUM 3
Страница 1: ...Instructions for Use QUEST SPECTRUM 3 ...
Страница 19: ...QUEST SPECTRUM 3 p 16 50 Instructions for Use QUEST SPECTRUM IPC Trolley Barco Branded 4K UHD surgical display ...
Страница 20: ...QUEST SPECTRUM 3 p 17 50 Instructions for Use Instructions for Use Draping Instructions Quick Start ...
Страница 26: ...QUEST SPECTRUM 3 p 23 50 Instructions for Use Figure 6 home screen Figure 7 insert patient data ...
Страница 27: ...QUEST SPECTRUM 3 p 24 50 Instructions for Use Figure 8 Start Surgery Figure 9 Select surgery ...