Operating Procedures
QIAcube User Manual 06/2008
5-13
A
B
Correctly loaded rotor adapter.
A
Rotor adapter is correctly loaded
and the 1.5
ml microcentrifuge tube lid is in the correct position;
B
Correctly loaded rotor adapter seen from the side.
A
B
Incorrectly loaded rotor adapter.
A
Rotor adapter is incorrectly
loaded with a 1.5
ml microcentrifuge tube. The lid of the tube is not
pushed all the way down to the bottom of the slot of the rotor adapter
and could break off during centrifugation, causing the protocol run to
crash (compare with part
A
of the previous figure
);
B
Incorrectly loaded
rotor adapter seen from the side (compare with part
B
of the previous
figure).
1.5 ml
micro-
centrifuge
tube lid is in
correct
Microcentrifuge
tube lid not
pushed down into
slot
Содержание QIAcube
Страница 1: ...Version 1 1 June 2008 Sample Assay Technologies QIAcube User Manual...
Страница 40: ...Introduction 2 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 43: ...General Description QIAcube User Manual 06 2008 3 3...
Страница 118: ...Maintenance Procedures 6 20 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 140: ...Troubleshooting 7 22 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 150: ...Appendix B B 2 QIAcube User Manual 06 2008 Centrifuge Shaker...
Страница 151: ...Appendix B QIAcube User Manual 06 2008 B 3 Centrifuge Shaker...
Страница 152: ...Appendix B B 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 156: ...Appendix C C 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 158: ...Appendix D D 2 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 161: ......