Operating Procedures
QIAcube User Manual 06/2008
5-19
Important
: Because the QIAcube cannot perform a load
check on these positions, it is absolutely necessary to use the
correct tube type filled with exactly the volume of buffer
described in the protocol sheet. Incorrect buffer volumes can
lead to pipetting errors during the protocol run. Make sure to
open the tubes before placing them onto the worktable. Lids
of microcentrifuge tubes should be securely placed into the
slots at positions A and C. If 3 tubes are to be placed onto
the QIAcube worktable, the lid of 1 tube must be cut off.
Correct loading of accessory buffers.
Incorrect loading of accessory buffers.
Lids of 2 of the microcentrifuge
tubes must be securely placed
in slots at positions A and C.
The lid of the
third tube
has not been
cut off, which
may lead to
a collision
with the tip
adapter.
The lid of the third tube must
be cut off.
Содержание QIAcube
Страница 1: ...Version 1 1 June 2008 Sample Assay Technologies QIAcube User Manual...
Страница 40: ...Introduction 2 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 43: ...General Description QIAcube User Manual 06 2008 3 3...
Страница 118: ...Maintenance Procedures 6 20 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 140: ...Troubleshooting 7 22 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 150: ...Appendix B B 2 QIAcube User Manual 06 2008 Centrifuge Shaker...
Страница 151: ...Appendix B QIAcube User Manual 06 2008 B 3 Centrifuge Shaker...
Страница 152: ...Appendix B B 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 156: ...Appendix C C 4 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 158: ...Appendix D D 2 QIAcube User Manual 06 2008 Page left intentionally blank...
Страница 161: ......