Maintenance, cleaning, servicing, disposal
CAUTION
The devices are subject to an annual safety and functional test (maintenance) in
accordance with §7 of the Medical Devices Operator Ordinance "Maintenance of
Medical Devices".
With the optional "smart assist ISO" accessory, a metrological test (MTP) is performed
at 2-year intervals in accordance with the manufacturer's specifications.
A selection of suitable products for maintenance can be found at
WARNING
The safety and effectiveness of the equipment can be optimally maintained if the
care and maintenance intervals are observed.
Component
Maintenance
interval
Testing and maintenance
Display
daily
Only use agents suitable for the display for cleaning and
disinfecting. Do not apply the agent directly, but use a
clean microfiber cloth.
Cushion
after training
Only use the disinfectants approved by the
manufacturer for disinfecting the pads.
A list of possible disinfectants can be found at
https://www.servicehelper.de/maintenance/.
Cladding
Floormats
Frame parts
regular
Clean with warm water and a microfiber cloth. Do not
use solvents, chlorides, polishes, chemical cleaners or
wax polishes.
Screws
monthly
Check externally accessible screws for tightness.
Guide rods
monthly
Check for sufficient lubrication and lubricate with suitable
maintenance oil.