Calibration by Proteus
To obtain a price quotation and a Return Material Authorization number for recalibration of your flow
meter, contact
When received at Proteus, your flow meter will:
1. Have its output measured and recorded in the as-received state.
2. The rotor, stainless steel shaft and sealing O-ring will be replaced.
3. The flow cavity will be cleaned and the device reassembled.
4. The unit will be recalibrated to its original specification against reference standards whose
calibrations are statistically controlled against NIST-traceable standards.
5. A new calibration certificate will be issued.
6. A new calibration label will be attached to the flow meter.
Direct calibration against NIST-traceable standards is optionally available.
Calibration by another laboratory
The calibrating laboratory will issue certificates and labels identifying the calibration status of your
metering flow switch.
Please advise your selected calibration laboratory to contact Proteus for calibration instructions.
Nylon, Real-Tuff and Hercules are trademarks of their respective holders
800 Series electronics are sensitive to Electro Static Discharge. There is normally no need to open the electronics case, but if it is
opened, proper ESD precautions should be taken.