Use Teflon tape or a PTFE-based liquid sealant to provide lubrication for the junction and a leak-tight 
connection at both input and output connections.  Real-Tuff and Hercules are two of many suitable 
brands of PTFE-based sealants.  
 

 

Do not over-tighten the connection. Refer to instructions for installation of the mating fittings for 
information on torque requirements. 
 

 

Leak testing of all connections in your flow circuit is recommended.  Pressurizing the system with air 
and external testing with a dilute soap solution can help identify leaking connections. 

 

 
SAE straight thread connections  

 
With these connectors, an O-ring makes the seal while the threads hold the connecting assembly in 
place.  Straight thread connections should receive a small amount of high-pressure lubricant before 
installation to prevent galling.   
 
 

Non-adjustable fittings 

 
 

1.  Bring the non-adjustable fitting into firm 

contact with the face of the port, using a 
wrench. 
 

2.  Check to be certain that the O-ring fits 

easily into the non-threaded receiving 
area of the port, and is not pinched. 

 

 
 

 

 

Adjustable fittings 

 

1.  Ensure that the locknut is positioned so 

the back-up washer is in contact with 
the beginning of the threads farthest 
from the end of the fitting. 
 

2.  Screw the fitting into the port until the 

back-up washer contacts the sealing 
face. 
 

3.  Check to be certain that the O-ring fits 

easily into the non-threaded receiving 
area of the port, and is not pinched or 
damaged. 
 

4.  Unscrew the fitting a maximum of one 

turn to position it in the desired direction. 
 

5.  Tighten the locknut firmly against the 

back-up washer so the fitting assembly 
is held securely in place. 

 

 
 
 
 
 
 

Filtering 

 
Your circulating fluid may contain particles.  While not essential to the operation of the flow sensor, it is 
good practice to filter your fluid.  A 100-micron filter is often used to remove rust and other particles from 
the fluid.  This can increase the lifetime of pumps and other fluid system components as well as reducing 
wear in the sensor. 
 

Содержание 0804BN03

Страница 1: ...800 Series Metering Flow Switches TECHNICAL REFERENCE MANUAL...

Страница 2: ...2 3 How the flow sensor works 7 4 Specifications and performance 8 5 Product certifications 9 6 Installation 9 Mechanical 9 Electrical 13 7 Selecting Trip Points 14 8 Measuring Flow Rate 16 9 Digital...

Страница 3: ...ince 1978 Section 2 Features and Functions Flow ranges from 0 05 to 50 GPM 0 2 to 190 LPM Eight flow ranges are provided in the 800 Series in brass polypropylene or stainless steel NPT and SAE connect...

Страница 4: ...ely retrofittable in place of existing Proteus flow switches The distance between the 800 Series pipe or fitting connections is identical to those of previous Proteus flow sensors Stainless Steel face...

Страница 5: ...nsor may be HIGHER than is suggested in Table 1 Model Number Switch Position 0 1 2 3 4 5 6 7 8 9 A B C D E F 0804BN03 0804PN03 0804SN03 0 01 0 03 0 05 0 06 0 08 0 10 0 12 0 14 0 16 0 17 0 19 0 21 0 23...

Страница 6: ...x flow rate at selected trip point 10 Actual Flow 11 5 GPM Closed Open Red Less than flow rate at selected trip point Actual Flow 10 GPM Open Closed Multiple outputs provide control flexibility Elect...

Страница 7: ...oltage is continuously compared to a user selected trip point voltage When the measured voltage is above the trip point the built in relay remains in its active state If the measured voltage falls bel...

Страница 8: ...tainless Steel Sealing O ring Viton Other materials are available Rotor Carbon fiber filled Nylon Rotor Shaft 316 Stainless Steel Other materials are available Meter Performance Voltage Output 0 5 VDC...

Страница 9: ...ing or tubing is used the 800 Series metering flow switch may be supported by direct connection to the pipe or tubing Panel mounting To mount the sensor behind a panel two of the faceplate securing sc...

Страница 10: ...PTFE tape or PTFE based liquid sealants to provide leak tight and lubricated junctions at all connection points Real Tuff and Hercules are two of many suitable brands of PTFE based sealants Note It is...

Страница 11: ...more than 10 x pipe ID be used between the flow sensor and any up stream devices to minimize these effects Appropriate calibration procedures must be used to provide an accurate flow measurement with...

Страница 12: ...ontact with the face of the port using a wrench 2 Check to be certain that the O ring fits easily into the non threaded receiving area of the port and is not pinched Adjustable fittings 1 Ensure that...

Страница 13: ...ith the electrical circuit and control functions of the system in which the sensor is to be included should perform installation of this product Electrical inputs and outputs are provided at a single...

Страница 14: ...necessary use the screwdriver to turn the fine adjustment potentiometer so that the indicator is pointed towards the 12 o clock position 6 Peel away the protective label covering the 16 step switch a...

Страница 15: ...flow Section 9 Digital Displays Digital display units Digital panel display meters are available to provide a direct local display of flow rate in selected engineering units Each unit has a 3 digit LC...

Страница 16: ...you connect to the proper terminals Connection instructions 1 Strip connecting wire leaving approximately bare wire exposed Stranded wire should be tinned with solder 2 Insert the wire into the screw...

Страница 17: ...ugh the flow meter In most cases annual cleaning immediately prior to recalibration is sufficient Tools required Wrenches to disconnect the flow meter from your flow circuit A flat screw driver Soft c...

Страница 18: ...surface of the rotor If the bearing surface is worn or not round replace the rotor Inspect the stainless steel shaft If the shaft shows signs of scoring or other wear replace the shaft or the whole fa...

Страница 19: ...er the calibrated response of the sensor Please contact Proteus Technical Support if a specialized calibration is needed Note As described in the Plumbing Connections section on page 10 the response o...

Страница 20: ...bel will be attached to the flow meter Direct calibration against NIST traceable standards is optionally available Calibration by another laboratory The calibrating laboratory will issue certificates...

Отзывы: