9.2.2 Cleaning after dismantling the UV lamp protection tube
WARNING!
UV-C radiation
Possible consequence: Serious injuries.
UV-C radiation is harmful to the eyes and skin
– Only start up the UV lamp
UV-C
when it is prop‐
erly installed
– Install the UV lamp into the UV system in
accordance with the regulations prior to com‐
missioning
WARNING!
Live parts!
Possible consequence: Fatal or very serious inju‐
ries
– Measure: Disconnect the device from the
power supply before opening it
– Disconnect damaged, defective devices or
devices that have been tampered with from the
power supply
CAUTION!
Fingerprints on the UV lamp
Possible consequence: Premature failure of the
UV lamp
– Only touch the glass of the UV lamp with
cotton gloves
– Fingerprints or impurities burn into the glass
and can result in premature failure
– For this reason always clean the lamp thor‐
oughly with a cloth moistened with alcohol
before installing it
– Then wipe the UV lamp with a dry, soft cloth
– Also thoroughly clean the glass of the UV lamp
return cable
Cleaning the UV sensor
Clean the UV sensor every time you clean the UV
lamp protection tube.
Safety switch for the UV lamp cover
The UV lamp cover is protected with a safety
switch. The safety switch has a reed switch and
comprises two components. The safely switch
ensures that the UV lamp only generates UV radia‐
tion in its fitted state. If the safety switch is
removed, then the UV lamp switches off. If the
safety switch is not fitted and thus remains closed,
e.g. after installation and maintenance work, then
the UV lamp cannot be started.