WARNING!
Consequence: Serious injuries.
UV-C radiation is harmful to the eyes and skin
– Only start up the UV lamp
UV-C
when it is prop‐
erly installed
– Install the UV lamp into the UV system in
accordance with the regulations prior to com‐
missioning
CAUTION!
Fingerprints on the UV lamp
Possible consequence: Premature failure of the
UV lamp
– Only touch the glass of the UV lamp when
wearing cotton gloves
– Fingerprints or impurities burn into the glass
and can result in premature failure
– For this reason always clean the lamp thor‐
oughly with a cloth moistened with alcohol
before installing it
– Then wipe the UV lamp with a soft, dry cloth
– Also thoroughly clean the glass of the UV lamp
return cable
1.
Use the START/STOP key to switch off the UV system.
Switch the main switch off or disconnect the mains plug from
the power supply.
2.
Close the shut-off valves upstream and downstream of the
radiation chamber.
Dismantling
Maintenance
71
Содержание ProMaqua Dulcodes A
Страница 82: ...A1903 Fig 79 Bearing cover 26 Open the bearing cover to obtain access to the slide bearing Maintenance 82...
Страница 118: ...118...