6.4.4 Functions of the keys in
[Startup] status
When the UV system starts up, the UV lamps are
first ignited and then the system waits until the UV
lamps have reached their operating temperature
and are able to provide their full UV power. If the
UV lamps have not ignited after 5 minutes, UV
system enters [Fault] status.
A1185
24/12/2012
10:06:53
STOP
MENU
IGNITING AND WARM UP LAMP
Lamp is starting
WIPER
FAULTS
Fig. 14: The UV system display in [Startup] status
Tab. 6: The following functions can be selected:
Key
Function
Upon selection of the function and subsequent pressing of Enter, the following
occur:
[F1]
[STOP]
The UV system is switched off
[F3]
[Wiper]
A wiper cycle is triggered
[F4]
[Fault]
Display of warnings and errors
[F5]
[Menu]
Change to the
[Menu] display
[F2,
F6 ...
F10]
- - -
- - -
Description
Function
[ESC]
- - -
[ACK]
Reset the fault
[TAB]
Selection of display fields
[DEL]
- - -
[HELP]
Operating tips
[ENTER]
Entry
Arrow keys
Selection of display fields
Control
27