Once you have selected the sensor type, the prompt automatically
appears asking whether the sensor parameters (zero point) have
to be determined. You can initiate this prompt manually as follows:
Continuous display ➨ Menu ➨ or
[Measurement] ➨ ➨
or
[Measuring channel X Conductivity] ➨ or
[Sensor parameter adjustment] ➨ .
1.
Use the arrow keys to select
[Automatically determine sensor parameters].
2.
Continue with .
ð
You will see the display showing
[Sensor dry] and
[Automatically determine sensor parameters].
3.
Continue with .
ð
You will see the display with the message
[Sensor parameters are automatically determined].
The sensor parameters are automatically carried over.
15.1.8.2
Calibrating conductive conductivity, cell constant
Prerequisite for calibration. The conductivity sensor is connected.
The conductivity sensor is in a conductivity calibration solution, the
conductivity of which is known.
Material
Order number
Conductivity calibration solution, 1413 μS/cm, 250 ml.
1027655
Conductivity calibration solution, 1413 μS/cm, 1000 ml.
1027656
Conductivity calibration solution, 12.88 mS/cm, 250 ml.
1027657
Conductivity calibration solution, 12.88 mS/cm, 1000 ml.
1027658
All parameters for the conductivity sensor are correctly entered in
the
[Measurement] menu item.
1.
Press in the continuous display.
2.
Use the arrow keys to select the channel which should be
calibrated.
3.
Then press
ð
You will see the display with the menu for selection of
[Cell constant] or [Temp. coefficient].
4.
Use the arrow keys to select
[Cell constant].
5.
Continue with .
ð
You will see the current data for the
[Cell constant]. Enter
the temperature coefficient of the calibration solution
here.
6.
Continue with calibration with .
7.
Continue with .
8.
Enter the known conductance of your conductivity calibration
solution.
9.
Accept with .
10. Continue with .
Calibration of the cell constant
Calibrating the controller diaLog DACb