User Aborts Protocol
If the protocol is aborted by the user, the Protocol Selected window will change to indicate that the protocol has
been aborted (Figure 48). After a protocol is aborted, the Current Step will be listed as Aborted by user. After
aborting the protocol, you can press the
Open Door
button to bring up the Clean Up Window (Figure 50).
An aborted run (initiated by the user or due to instrument error) will result in the loss of all the
samples.
Do not attempt to repurify samples from an aborted run.
Figure 48. Protocol Selected window after protocol abort by the user.
If a protocol is aborted by
the user, the Protocol Selected window changes to display a red title bar and run progress bar. The Current
Step changes to Aborted by user. After a protocol is aborted you can touch the
Open Door
button to run
the Clean Up procedure.
An aborted run (initiated by the user or due to instrument error) will result in the loss of all
the samples.
Do not attempt to repurify samples from an aborted run.
!