4.F. Maxwell
®
RSC User Interface Screens (continued)
Self Test
You can run a ‘Self Test’ by touching the
Self Test
button from the Maxwell
®
RSC Settings screen
(Figure 31). On touching the
Self Test
button, the Maxwell
®
RSC Instrument will perform a routine test
(Figure 31) to confirm that instrument functions including initialization of the deck tray, plunger bar and
magnet bar, motion of these systems, and the instrument heater are operating within acceptable perfor-
mance ranges. Performing the Self Test will result in the generation of a system report that details the Pass/
Fail status of the tests that are performed. Following the Self Test the software automatically opens this
report.
Figure 30. Maxwell
®
RSC Settings screen.
From this screen you can perform Self Test for the
Maxwell
®
RSC Instrument.
Figure 31. Self Test protocol being executed.