4.F. Maxwell
®
RSC User Interface Screens (continued)
Options
Administrators can view additional options available in the RSC software by touching the
Options
button
on the Administrator page window. The Options window displays a set of three options that can be used by
the administrator to adapt the functionality of the software to the needs of the laboratory (Figure 29). By
default all of the options below are unchecked. Below are listed the options that can be set and a description
of their functions.
•
Allow use of expired kits:
Checking this box allows users to run methods with kits that have passed
their expiration date. When this box is unchecked the software will not allow expired kits to be used for
any purification runs.
•
Allow deletion of results:
When checked, this box will allow deletion of extraction report files from
the database of completed run reports.
•
Save aborted runs:
When this box is checked, the software will save run reports for all instrument
functions regardless of why they were aborted. If left unchecked, any instrument functions that have
been aborted will not generate a run report.
Figure 27. Methods window.
This window displays a list of the methods that are currently installed in
the Maxwell
®
RSC User Interface. Touching the
Select Import Package
displays a file path for navigating
to a particular file (Figure 28). This file will be used to import a new or updated method.