17.D. UV-Vis Absorbance Module
To add UV-Vis detection capability to the GloMax
®
-Multi+ Detection System,
contact Promega Customer Services or a Distributor for information on the
purchase and installation of the UV-Vis Detection Module. See Section 22.D,
Appendix, for ordering information.
Installing the UV-Vis Absorbance Housing Unit
1. Touch the “Door” button to automatically close the instrument door and
reset the Microplate Sample Tray to Home position.
2. Power off the GloMax
®
-Multi+ Detection System and disconnect the power
supply from the back of the instrument.
3. Gently tip the GloMax
®
-Multi+ Detection System so that it rests on its back
panel so that the bottom of the instrument is accessible.
Figure 51. Removing the metal cover plate.
4. Using a 3/16 inch Allen wrench remove the four 3/16 inch Allen head
screws from the metal plate cover exposing the interior compartment
(Figure 51).
5. Carefully insert the connector on UV-Vis housing unit into the connector
inside the bottom of the GloMax
®
-Multi+ Detection System.
Figure 52. Attaching the UV-Vis housing unit.
7960T
A
7959T
A