Figure 48. Installing the Absorbance Module.
5. Slide the Absorbance Module into position. Be cautious to not damage the
printed circuit board on the back of the detection head.
6. Hand tighten the two captive screws using the Allen wrench (size 7/16 inches)
provided with the kit. Too much force may damage the screws. See
Figure 49.
Figure 49. Arrows point to two captive screws on the Absorbance Module.
7. Close the instrument door.
8. Power on the instrument.
9. To ensure the instrument detects the newly installed module, verify the
“Absorbance” icon appears on the Instrument Status bar on the Instrument
Control screen. See Figure 50.
Figure 50. Detection Mode icon on the Instrument Status bar indicating the
Absorbance Module is detected.