5. Carefully insert the connector on the Visible Absorbance housing unit into
the connector inside the bottom of the GloMax
®
-Multi+ Detection System
(Figure 47).
Figure 47. Attaching the Visible Absorbance housing unit.
6. Using the provided 3/16 inch Allen wrench, tighten the four 3/4 inch
Allen head screws.
7. Manually open the instrument door and install the Visible Absorbance
Module.
8. Power on the instrument.
9. To ensure that the instrument is detecting the newly installed Visible
Absorbance Module, verify that the “Absorbance” icon appears on the
Instrument Status bar of the Instrument Control screen.
Installing the Absorbance Module
1. If the door is open, touch the “Door” button to automatically close the
instrument door and set the Microplate Sample Tray into Home position.
2. Turn off the GloMax
®
-Multi+ Detection System.
3. Manually open the instrument door, and hold it open.
4. Hold the Absorbance Module with the fingertips of one hand. Guide the
Absorbance Module into the threaded stainless steel post located on the left
side of the instrument interior. See Figure 48.