Page 77
Figure 67. Replacing the Microplate Sample Tray Break-Away tab.
3. Touch the “Door” button to automatically close the Microplate Sample
Tray Cover and retract the Microplate Tray Assembly. Do not push down
on the Microplate Sample Cover or force it to close.
Replacing the Microplate Sample Tray Cover Break-Away Tab
If force has been used on the Microplate Sample Tray Cover, the Break-Away
tab will break, preventing damage to the tray assembly. If the Break-Away tab
is broken, the Tray Cover will not stay upright.
1. Touch the “Door” button to automatically close the instrument door and
set the Microplate Sample Tray into Home position.
2. Turn off power to the GloMax
®
-Multi+ Instrument.
3. Manually open the instrument door and hold it open.
4. Use a Phillips head screw driver to remove the broken Break-Away tab and
replace it with a new tab, included with the accessories.
5. Touch the “Door” button to automatically close the Microplate Sample
Tray Cover and retract the tray assembly back into the instrument. Do not
push down on the Microplate Sample Plate Cover or force it to close.