Figure 68. Injector tip assembly.
20.G. Replacing Injector Tips
Injector tip replacements are available. Only Promega-authorized injector tips
are fully tested and compatible with the GloMax
®
-Multi+ Detection System.
See Section 22.D for ordering information.
1. Manually open the instrument door, and hold it down with one hand.
2. Gently remove the injector tip assembly from the injector tip holder located
on the Luminescence Module.
3. Twist the white tip valve counterclockwise to release it from the gray tube
fitting. See Figure 68 for a picture of the (white) tip valve.
4. Untwist the white tip fitting from the white valve to release the tip.
5. To remove the tip, hold the tip at the base and twist off.
6. Discard tip.
7. Insert a new tip onto the valve.
8. To complete the injector tip assembly, twist the fitting clockwise until tight.
A small gap of approximately 1mm between the two fittings is normal.
9. Twist the assembled fitting with the new tip clockwise onto the tube fitting.
10. Insert the tip into the injector tip holder on the Luminescence Module.