STRETCHER SUPPORT SERIES PB460 AND PB455
Review 2022/02
25
03.5 Troubleshooting
PROBLEM
CAUSE
SOLUTION
Structural damage
Improper use or operators not properly trained
Take the stretcher support out of service imme-
diately and contact the technical service
The lock lever does not return to
its position.
The spring is damaged
Replace spring.
If the problem persists, contact the technical
service.
When operationg the levers the
controls the platform does not
extract
The cables are not tight enough
Tighten the cables with the tensioners.
If the problem persists, contact the technical
service.
When activating the pedal,
the platform cannot be moved
laterally.
The cables are not tight enough
Tighten the cables with the tensioners.
If the problem persists, contact the technical
service.
When operating the controls, the
bed does not tilt
The microswitches are poorly positioned
Reposition the microswitches so that they make
contact with the lever stop.
If the problem persists, contact the technical
service.