OPERATION
- The state of the springs
- That the welds are intact, without cracks or breaks.
- That there are no tubes or metal sheets that have bends or cracks.
- Make sure that the vehicle in which you want to install the device is manufactured in accordance with EN
1789 standard.
- The device is supplied in an intermediate position to facilitate installation. Do not connect the device to
the power source before it has been fixed to the ground.
If the above conditions are met, the device can be considered ready for use; otherwise, it should be imme-
diately removed from service and contact the manufacturer.
03.2.1 ANCHOR POINTS
Before installing the
stretcher support
, check that the ambulance is capable of storing it, allowing it to move
in all directions.
Make sure that the installation floor is perfectly flat. Non-flat surfaces or parts of the floor that are not per-
fectly level could compromise the correct operation of the product.
Before putting the base into operation, you must ensure that each and every one of the anchor points is in
place and tightened.