03.2.2 CONNECTION TO THE ELECTRIC GRID
To make the electrical connection of the stretcher support in the ambulance, follow these steps:
1. Locate the black cable placed on the front of the stretcher support (according to the direction of the
ambulance)
2. Remove the plastic end of the red and black wires with a wire stripper to allow connection.
3. Use a power strip and a flathead screwdriver to connect the red and black wires with the ambulance
outlet.
The voltage required for the stretcher support is 12V in direct current.
Do not leave the cables in walkways or with moving elements, it could cause breakages, cuts and failures
in the system.
ATTENTION: Make sure you have the power line properly protected and the power along
the line is stable.
REST POSITION:
It is recommended the rest position to be halfway of both motors.
This way, when you want to place the stretcher support in any other position, the two motors will already
be halfway whether they have to close or open completely.
TRENDELENBURG POSITION:
It is only necessary to adjust the Trendelenburg position in the event that the stretcher support has a su-
pplement that causes interference with the ground.
Otherwise, the stretcher support is factory set to the maximum Trendelenburg position.
When finished, store the programming remote in a separate location from the control remote to avoid con-
fusion. When the user’s hand control is used for the first time after programming, the bed will go to the rest
position and then lower to its minimum position.
5 cm
FIGURE 3
The stretcher support must be programmed in the fo-
llowing positions:
LOAD POSITION:
To configure the loading height of the stretcher support,
the optimal position of the stretcher for loading and un-
loading must be considered.
The position is as follows: when the front wheels of the
stretcher are placed on the platform and the front legs
touch the ramp roller, the front wheels of the stretcher
must be at least 5 cm from the ground (see figure 3).
03.2.3 STRETCHER SUPPORT PROGRAMMING
The stretcher support is delivered with two hand controls: for programming and for users. The following
indicates how to program the bench with the programming and control (see figure 4).
ATTENTION: A poor adjustment of the loading position of the stretcher support can lead
to a malfunction of opening and closing of the legs of the stretcher.