July 2018
6-75
6
SPECIFICATIONS
6.1
Specifications AE-366B
FREQUENCY
Frequency Range
Setting Range
150 kHz to 3 GHz.
Center Frequency
Setting Resolution
0.1 MHz.
Accuracy
within
±
50kHz
(frequency span: 0.3 GHz to 2.6 GHz, 20
±
5
°
C).
Frequency Span
Setting Range
1 MHz to 3 GHz.
Accuracy
within
±
3%
(frequency span: 0.3 GHz to 2.6 GHz, 20
±
5
°
C).
Resolution Bandwidth
Setting Range
30 kHz, 100 kHz, 300 kHz, 1 MHz.
SSB Phase Noise
-85 dBc/Hz (typical, 500 kHz offset, RBW: 30 kHz, Sweep time: 1.5s,
Span:1MHz@1GHz)
Inherent Spurious Response
less than –45dBc @ -40 dBm Ref. Level (typical less than –50 dBc).
AMPLITUDE
Reference Level
Input Range
+20 to –40 dBm.
Unit
dBm, dBV, dB
μ
V.
Average Noise Level
≤
-100 dBm.
(typical, center frequency:1 GHz RBW:30 kHz)
Frequency Characteristic
within
±
3.0dB @300MHz
∼
2.6 GHz
within
±
6.0dB @ 80
∼
300 MHz, 2.6
∼
3 GHz.
Accuracy
within
±
2 dB (1 GHz); SPAN: 5 MHz; Ref. Level odBm, input signal
–10 dBm.
Input
Input impedance
50
Ω
.
Input VSWR
less than 2.0@input att
≥
10 dB.
Input damage level
+30 dBm (CW average power), 25 VDC.
Input connector
N
connector.
Содержание AE-366B
Страница 1: ...AE 366B SPECTRUM ANALYSER 0 MI2007...
Страница 8: ......
Страница 57: ...July 2018 3 49 Figure 48 Average Off Figure 49 Average On 8x...