July 2018
3-19
3.2.3
Zero Span
The Zero Span function will set the frequency range of the sweep to 0 Hz and
fixes the start and stop frequencies to the center frequency.
The Zero Span function measures the time domain characteristics of the input
signal at the center frequency. The horizontal axis is displayed in the time
domain.
Press
X
Zero Span
.
The span changes accordingly.
Figure 20.
Amplitude modulation.
NOTE
:
The measurement functions such as ACPR and OCBW are not available
with the zero span setting.
3.2.4
Last Span
The last span function returns the spectrum analyzer to the previous span
settings.
Press
X
Last Span
.