July 2018
3-53
3.11
Save/Recall Files
The
AE-366B
can save and recall setup data, trace data and limit line data to
and from internal memory. There are five memory locations for each save file
type. These files cannot be saved to USB.
The Hardcopy key can be used to save image files to a USB flash drive.
3.11.1
Save/Recall Setup
Setup data contains all the data necessary to recall the state of the
AE-366B
to
known state.
Setup data contains the following data:
Center frequency, Start frequency, Stop frequency, Step frequency, Ref. Level,
Scale, Units, RBW.
Save
:
To save the current settings, press
X
Setup To
and
choose a memory location to save to with the arrow keys.
Setup To:
1
~
5.
Press
to execute the save.
Recall
:
To recall a setup, press
X
Setup From
and choose a
memory location to recall from with the arrow keys.
Setup From:
1
~
5.
Press
to execute the recall.