5
FAQ
►
I connected the signal but it does not appear on screen.
►
The trace is updated too slowly.
►
I cannot see the trace on the screen.
►
The performance does not match the specification.
I connected the signal but it does not appear on screen.
Run Autoset and let the
AE-366B
find the best display scale for your
target signal. Press the Autoset key, then press Autoset
.
The trace is updated too slowly on the screen.
The sweep time determines how often the trace is updated on the screen.
To increase the sweep time, try reducing the span or using a wider RBW
setting.
I cannot see the trace on the screen.
If you cannot see the trace on the screen, there may be a number of
possibilities.
The trace is just off screen
: Try to adjust the reference level with
the amplitude key.
The trace may be in the “Blank” mode: Putting the trace into view
mode will enable the trace to be viewed again.
The performance does not match the specification.
Make sure the device is powered On for at least 30 minutes, within
+20 °C ~ +30 °C. This is necessary to stabilize the unit to match the
specification.
5-74
July 2018