8. Technical Specifications
Maximum Tensile Force
Maximum Pulling Speed
Proceq DY-206
6 kN
4.65 mm/min
0.183 inch/min
1349 lbf
Proceq DY-216
16 kN
3597 lbf
Proceq DY-225
25 kN
2.2 mm/min
0.086 inch/min
5620 lbf
Max stroke
5 mm
Calibration accuracy
EN ISO 7500-1 Class 1 (±1% from 20% of max. Force)
Memory capacity
100 measurements
Battery capacity
1500 mAh, 3.7V (min. 80 measurements)
Charger connection
USB type A (5V, 500mA)
Weight
4.5 kg
Dimensions of housing
109 x 240 x 205.5 mm
Operating temperature
-10 to 50°C (14 to 122°F)
Storage temperature
-10 to 70°C (14 to 158°F)
IP classification
IP54
8.1 Maintenance and Support
Maintenance
To ensure reliable and accurate operation, it is recommended to have the instrument serviced and
re-calibrated on a yearly basis. Customers may however determine their own service interval based
on application and usage. The instrument assists with determining a suitable interval by giving a
warning on start-up after 1500 complete test cycles.
Support Concept
Proceq is committed to providing a complete support service for this instrument by means of our
global service and support facilities. It is recommended that the user register the product on www.
proceq.com to obtain the latest on available updates and other valuable information.
8.2 Standard Warranty and Extended Warranty
The standard warranty covers the electronic portion of the instrument for 24 months and the me-
chanical portion of the instrument for 6 months. An extended warranty for one, two or three years