background image

12 

 

 © 2014 Proceq SA

5.3  Failure Mode Recording

Many standards also require the user to record the mode of failure in varying degrees of complex-
ity. Proceq DY-2 allows the user to record the failure mode with the test result for comprehensive 
reporting. The possibility to record the failure is flexible enough to accommodate the requirements 
of the various standards.
e.g. ISO 4624 failure mode reporting classification.
•  A is a cohesive failure of the substrate;

•  A/B is an adhesive failure between the substrate and the first coat;

•  B is a cohesive failure of the first coat;

•  B/C is an adhesive failure between the first and second coats;

•  etc.

•  Estimate the area of fracture as a percentage to the nearest 10% for each type of fracture. 

The first column indicates where the break took place. The second column indicates the percentage 
of the break in that particular layer.
For simplified reporting as specified for example in ASTM C 1583, then the user would enter for 
example B 100% indicating a failure at the first coating.
•  Use the navigation keys to select the item from either column.

•  Define your own classifications for A, B, etc.

•  Use the center key to alter the value.

5.4  Emergency Stop

In case of an emergency the red-colored STOP button can be pressed. The instrument stops in any 
case and the following dialog appears:

•  Selecting the "Rewind" icon lets the instrument move back to its starting postition.

•  Selecting the "Cancel" icon lets you return to the main menu without any movement of the motor 

and you may proceed with the test.

Содержание DY-2 series

Страница 1: ...Proceq DY 2 Operating Instructions Swiss Precision since 1954...

Страница 2: ......

Страница 3: ...oceq DY 2 Menu 6 4 Test Parameter Settings 6 5 Performing a Pull Off Test with Proceq DY 2 9 5 1 Test Preparation 9 5 2 Measurement Procedure 10 5 2 1 Peak load test duration and effective load rate 1...

Страница 4: ...e check operation and maintenance of the instrument and its components Unauthorized structural modifications to the instrument and its components Serious damage resulting from the effects of foreign b...

Страница 5: ...th by near to surface tests 50mm 0 05 0 03 MPa s ASTM D 4541 Standard test method for pull off strength of coatings using portable adhesion testers 1 MPa s 100s ASTM C 1583 Standard test method for te...

Страница 6: ...ded for elastic surfaces due to higher requirement on load rate Proceq DY 216 covering most applications Recommended for elastic surfaces due to higher requirement on load rate Proceq DY 225 required...

Страница 7: ...nt A complete charging cycle will take approximately 3 hours longer in the case of a deep discharge Note When charging the battery or when operating on mains supply the charging circuit will be discon...

Страница 8: ...ts down automatically after 5 minutes of inactivity 3 3 The Proceq DY 2 Menu On start up the main menu screen will be shown Set Measurement Unit Start Test Memory Select Test Disc Diameter Set Test Di...

Страница 9: ...m value If no maximum load is required this should be set at the maximum value supported by the instrument i e 6kN 16kN or 25kN Press the center key to complete the action Load rate Set the load rate...

Страница 10: ...00 measurements When completely full the oldest will be overwritten To review the saved measurements follow the procedure below Choose the memory icon Choose the icon to review the measurements Scroll...

Страница 11: ...s are The size of the test disc The load rate applied to the test disc The total time allowed for the test 5 1 Test Preparation Pull off testing is used on many different types of materials The actual...

Страница 12: ...If the surface is inclined this method is not practical 5 2 Measurement Procedure Press the center key to begin Verify settings Note In the case of square discs the diameter is a theoretical value Pa...

Страница 13: ...at the graphical output in this case for confirmation The effective load rate is calculated from the point at which load rate control begins 250N to the time at which 80 of the peak load is reached T...

Страница 14: ...cture as a percentage to the nearest 10 for each type of fracture The first column indicates where the break took place The second column indicates the percentage of the break in that particular layer...

Страница 15: ...load must not exceed 16kN See Chapter 4 Set the Load Parameters 6 2 Working on vertical surfaces or overhead When working on vertical surfaces or overhead ensure that the test disc is firmly glued to...

Страница 16: ...ions calibration certificate and carrying case 7 2 Test Discs and Accessories 346 10 500S Test disc steel 50mm M10 set of 10 346 10 501S Test disc aluminum 50mm M10 set of 10 346 10 502S Test disc alu...

Страница 17: ...have the instrument serviced and re calibrated on a yearly basis Customers may however determine their own service interval based on application and usage The instrument assists with determining a su...

Страница 18: ...stored on Proceq DY 2 Double click on the DY Link icon on your desktop or start DY Link via the start menu DY Link starts with a blank list Application settings The menu item File Application setting...

Страница 19: ...ty programs Click on the measurement object s you wish to export Click on the Export measurements to CSV file s icon The data for this measurement object s is exported as a Microsoft Office Excel comm...

Страница 20: ...op of the screen PQUpgrade icon Allows you to upgrade your firmware via the internet or from local files Open DY link project file icon Allows you to open a previously saved pqr project Save project i...

Страница 21: ...window Start the test on the Proceq DY 2 and the load increase will be displayed in real time until the test is completed Note This curve is for monitoring purposes only You have to perform a download...

Страница 22: ...2014 Proceq SA 20 Notes...

Страница 23: ...2014 Proceq SA 21 Notes...

Страница 24: ...4 korp 2 lit A Office 410 197374 St Petersburg Russia Phone Fax 7 812 448 35 00 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharjah United Arab Emirates Phone 971 6 557 8505 Fax 9...

Отзывы: