Click on the “Live View” icon to open the dialog:
The measurement update rate determines how
many data points are used to build up the curve.
The Max Y-axis may be set either at automatic or
using the maximum load setting on the instrument.
The default setting is to use a single window to
show the load rate. If you wish to make compari-
sons between more than one test, then each subse-
quent test can be opened in its own window.
Press start to open the Live View
window:
Start the test on the Proceq DY-2 and the load
increase will be displayed in real-time until the test
is completed
Note: This curve is for monitoring purposes only. You have to perform a download of the
saved measurements from the Proceq DY-2 to see the saved curve data.
Calibration
Proceq Proceq DY-2 is calibrated according to EN ISO 7500-1 Annex C - Alternative Method of
Testing Machine Classification.
This software allows a re-calibration or a reset to an earlier saved calibration.
Typically this will only be performed by a service centre or calibration laboratory.
Separate Operating Instructions describing the calibration procedure can be found under the "Help"
menu.