The
video mode
only functions when using Ethernet communications. The HP-MSM-HB provides 4 frames
per second in
video mode
. Unlike the monitoring operation, only raw data is transmitted in
video mode
.
If the detector is over driven during a measurement (the appearance of red in the representation and/or an
A/D-Converter value of 4 095 in a clear section indicates
signal saturation
), you should reduce the amplifi-
cation using the Amplitude slider
Ampl.
and repeat the measurement. Amplification is automatically regulated
when optimization is activated. If necessary, filtering must also be increased using a neutral glass density
filter.
The radiated power can be configured using the
Power
scroll bar. The reference value for the scroll bar is
entered in the
Measurement > Environment
menu. The power density is calculated in relation to the power
values set here. Up to 50 individual measurements can be recorded in a measurement file. This is relevant
when measuring the beam caustic and for time or power series. It is possible to switch back and forth be-
tween the individual measuring planes.
With the button
Copy
the measurement settings (window size and position, power and amplification) can be
copied from the previous measuring plane.
By means of the option
Averaging
the average of the results of up to 50 single measurements per each
plane is determined. There are different analysis algorithms available:
•
Average
determines the average value of the distributions measured
•
Max. pixel
determines the point wise maxima of the distributions measured
•
Max. trace
determines the maximum traces of the distributions measured
During a measurement, the status of the measurement system is constantly displayed. These are:
• the current measuring plane
• the run of the reference cycle
• positioning the measuring head
• the measurement
• the data transmission – the progress is shown by means of the bar display
Fig. 13.10: Display window
Monitor
By means of the button
Stop
you can cancel a running measurement (this does not disrupt movement along
the z-axis). We recommend triggering a
Reset
cycle afterwards. Canceling will also end
monitor
operation.
The
Stop
button can be used to stop the measurement. A
Reset
cycle should then be triggered before
further measuring.
Содержание HP-MSM-HB
Страница 2: ......
Страница 3: ...3 Revision 00 EN 01 2021 HP MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 72: ...72 Revision 00 EN 01 2021 HP MSM HB 35 0 6 16 Accessories...
Страница 77: ...77 Revision 00 EN 01 2021 HP MSM HB 35 0 6 19 Declaration of conformity...
Страница 149: ...149 Revision 00 EN 01 2021 HP MSM HB 35 0 6...