![Primes HP-MSM-HB Скачать руководство пользователя страница 116](http://html1.mh-extra.com/html/primes/hp-msm-hb/hp-msm-hb_original-instructions-manual_1601727116.webp)
116
Revision 00 EN - 01/2021
HP-MSM-HB
35,0(6
1
Single
Monitor
Video Mode
Starts a measurement in the chosen plane
Starts repeated measurements in the chosen plane automatically
2
Start
Starts a measurement in the currently chosen plane
3
Stop
Finishes the measurement in the currently chosen plane
4
Reset
The measuring device is reset
5
Stop Motor
Not relevant for HP-MSM-HB
6
Plane
Selection of the measuring plane (0-49) either explicit or by means of the buttons (+/-)
7
Entry field
Numerical entry of the z-position
8
Copy
Copies all settings (window size and window position; x, y, z; etc.) from the former plane
to the current plane (e.g. 1>>2)
9
Find beam
Starts an automatic beam search in the current measuring plane
10
Scan
Starts an automatic beam search with the HP-MSM-HB. The algorithm works at a fixed
z-position and searches only within the range of the specified measuring window.
11
Ampl.
Slide control in order to adjust the optical amplification (exposure time of the CCD)
12
Power
Slide control in order to adjust the laser power to save it in the software
13
Entry field Power
Numerical input of the laser power to save it in the software
14
Entry field Ampl.
Numerical input of the electrical amplification
15
Averaging
Analysis of the serial measurements. Averaging algorithms: average value, values of the
maximum pixels and the value of the maximum trace
16
Averaging
Selectable number (1 – 50) of single measurements for the averaging
17
LED symbol and
bar graph display
Display for the degree of the signal saturation (LED green
≙
ok, red
≙
not ok)
18
False color
Activates the option of the false color presentation
19
Zoom
Magnification settings for the measuring window
20
Symmetric
This option enforces the usage of square measurement windows, whose size is only
adjustable via x.
21
X/Y
Set the measurement window size for non-square windows
22
Display
Measuring window shows the current measuring result
23
Z
Slide control in order to set the z-position
Tab. 24.1: Explanation of input and setting elements
Содержание HP-MSM-HB
Страница 2: ......
Страница 3: ...3 Revision 00 EN 01 2021 HP MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 72: ...72 Revision 00 EN 01 2021 HP MSM HB 35 0 6 16 Accessories...
Страница 77: ...77 Revision 00 EN 01 2021 HP MSM HB 35 0 6 19 Declaration of conformity...
Страница 149: ...149 Revision 00 EN 01 2021 HP MSM HB 35 0 6...