Flowchart of a measurement
13.5.1
Prepare measurement
Yes
ok
Too
low
Too high
No
Search caustic
settings
Select Magnification
Low x 0.3
on the device
Increase laser
power
Increase
filter
Set laser power on
minimum
Positioning z-axis:
HP-MSM-HB = 60 mm
HP-MSM-HB20kW= 20mm
1. mea-
surement
with largest
window
Beam
found?
Check
saturation
S/N
Preparation
Check alignment
again
Laser beam is
switched on?
OD filter is
selected?
13.5.2
Set caustic limits
No
Yes
Increase
z-axis value
Increase z-axis value
until the beam fits into
the biggest window
Upper limit value for
the caustic measure-
ment found
Decrease the z-axis
value until the beam fits
to the biggest window
Increase the z-axis value
until the beam fits to the
biggest window
Lower limit value for the
caustic measurement
found
Increase z-axis value
until the beam fits into
the biggest window
Upper limit value for the
caustic measurement
found
Decrease z-axis value
until the beam fits the
biggest window
Lower limit value for
the caustic measure-
ment found
Beam fits
into the
biggest win-
dow
Search caustic
settings
Start first
caustic measure-
ment
Beam gets smaller
Beam gets bigger
Содержание HP-MSM-HB
Страница 2: ......
Страница 3: ...3 Revision 00 EN 01 2021 HP MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 72: ...72 Revision 00 EN 01 2021 HP MSM HB 35 0 6 16 Accessories...
Страница 77: ...77 Revision 00 EN 01 2021 HP MSM HB 35 0 6 19 Declaration of conformity...
Страница 149: ...149 Revision 00 EN 01 2021 HP MSM HB 35 0 6...