![Primes HP-MSM-HB Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/primes/hp-msm-hb/hp-msm-hb_original-instructions-manual_1601727027.webp)
27
Revision 00 EN - 01/2021
HP-MSM-HB
35,0(6
8
Cooling water and compressed air connections
For the operation of the HP-MSM-HB both a cooling water and a compressed air supply is required.
The device is equipped with a HighBrilliance measuring objective, which has to be cooled with water and
flushed with compressed air. The required hoses are already preassembled.
Cooling Circuit (return flow)
Compressed Air Supply
(on the face side)
Cooling Circuit (inlet flow)
Fig. 8.1:
Media Connections of the HP-MSM-HB (20 kW)
NOTICE
Fire hazard; Damage/Destruction of the device due to overheating
If there is no water cooling or a water flow rate which is insufficient, there is a danger of
overheating, which can damage the device or set it on fire.
X
Operate the device with a connected water cooling and a sufficient water flow rate only.
8.1
Cooling Circuit System
For packaging reasons the cooling water pipe of the absorber was removed. The device must not be oper-
ated without water cooling. The removed cooling water pipe has to be reconnected with the absorber.
X
Remove the sealing plug from both the water
pipe and the absorber.
Содержание HP-MSM-HB
Страница 2: ......
Страница 3: ...3 Revision 00 EN 01 2021 HP MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 72: ...72 Revision 00 EN 01 2021 HP MSM HB 35 0 6 16 Accessories...
Страница 77: ...77 Revision 00 EN 01 2021 HP MSM HB 35 0 6 19 Declaration of conformity...
Страница 149: ...149 Revision 00 EN 01 2021 HP MSM HB 35 0 6...