™ PD7000 Temperature Meter 

Instruction Manual 

 

38 

Troubleshooting Tips 

This meter is a highly sophisticated instrument with an extensive list of features and capabilities. If the front panel 
buttons are used to program the meter, it may be a difficult task to keep everything straight. That is why we strongly 
recommend the use of the free 

MeterView Pro

 software for all programming activities. A cable is provided with the 

meter for programming with MeterView Pro software.   

If you have programmed the meter with the front panel buttons and it is not working as intended, try  
re-programming the meter using MeterView Pro software. 

Symptom 

Check/Action 

No display at all 

Check power at power connector 

Not able to change setup or 
programming, 

Locd

 is displayed 

Meter is password-protected, enter correct six-digit password to 
unlock 

Meter displays error message during 
calibration (

Error

Check: 
1.  Signal connections 
2.  Type and range selector switches 
3. 

Input selected in 

Setup

 menu 

Meter displays 
1. 

99999  

2. 

-9999

  

Check: 
1.  Input selected in 

Setup

 menu 

2.  Corresponding signal at Signal connector 

Display is unstable 

Check: 
1.  Input signal stability and value 
2.  Filter and bypass values (increase) 

Display response is too slow 

Check filter and bypass values 

Display reading is not accurate 

Check: 
1.  TC wire type 
2.  Input selection in 

Setup

 menu 

Display does not respond to input 
changes, reading a fixed number 

Check: 
1.  Display assignment, it might be displaying max, min, or set 

point. 

Display alternates between 

1. 

Hi

 and a number  

2. 

Lo

 and a number 

Press Menu to exit max/min display readings.  

Relay operation is reversed 

Check: 
1.  Fail-safe in 

Setup

 menu 

2.  Wiring of relay contacts 

Relay and status LED do not  
respond to signal 

Check: 
1.  Relay action in 

Setup

 menu 

2.  Set and reset points 

Flashing relay status LEDs 

Relays in manual control mode or relay interlock switches opened.  

Meter not communicating with 
application programs 

Check: 
1.  Serial adapter and cable 
2.  Serial settings 
3.  Meter address and baud rate 

If the display locks up or the meter does 
not respond at all  

Cycle the power to reboot the  
microprocessor. 

Other symptoms not described above 

Call Technical Support for  
assistance. 

Note: Certain sequences of events can cause unexpected results. To solve these issues, it is best to start fresh from 
factory defaults and use the manual as a step by step programming guide, rather than a random approach to 
programming. To reset the meter to factory defaults, see Reset Meter to Factory Defaults on page 36. In addition, for 
best results, we recommend using the free MeterView Pro software for all programming needs.

 

 
 

Содержание ProVu PD7000

Страница 1: ...with Interlocking Capability Isolated 4 20 mA Output Options Free PC Based Onboard MeterView Pro USB Programming Software No Assembly Required Optional SunBright Display Models for Outdoor Applicatio...

Страница 2: ...gistered Trademarks All trademarks mentioned in this document are the property of their respective owners 2020 Precision Digital Corporation All rights reserved FREE MeterView Pro Programming Software...

Страница 3: ...tions 15 Analog Output Power Supply 15 External Relays Digital I O Connections 16 Interlock Relay Feature 16 Setup and Programming 17 Front Panel Buttons and Status LED Indicators 17 Display Functions...

Страница 4: ...Making Changes to a Password Protected Meter 29 Disabling Password Protection 29 Advanced Features Menu 30 Advanced Features Menu Display Messages 30 Offset Adjust Adjust 31 Recalibration of the Mete...

Страница 5: ...1 Thermocouple Input Connections 14 Figure 12 Three Wire RTD Input Connections 14 Figure 13 Two Wire RTD Input Connections 14 Figure 14 Four Wire RTD Input Connections 14 Figure 15 Average Temperature...

Страница 6: ...to any system Ordering Information Standard Models 85 265 VAC Model 12 24 VDC Model Options Installed PD7000 6R0 PD7000 7R0 No options PD7000 6R2 PD7000 7R2 2 relays PD7000 6R3 PD7000 7R3 4 20 mA outp...

Страница 7: ...igital inputs Pass 2 Allows use of function keys digital inputs and editing set reset points Pass 3 Restricts all programming function keys and digital inputs Non Volatile Memory All programmed settin...

Страница 8: ...50 0 4 C 120 110 to 500 0 2 F 79 to 260 0 1 C 1000 328 to 900 0 7 F 200 to 482 0 4 C All ranges capable of 1 or 0 1 resolution Relays Rating 2 or 4 SPDT Form C internal and or 4 SPST Form A external r...

Страница 9: ...wer USB port provides power to the meter DO NOT apply AC or DC power to the meter while the USB port is in use On Board Digital Input F4 Function Remote operation of front panel buttons acknowledge re...

Страница 10: ...10V 0 15 80 MHz 1kHz 80 AM AC Surge 2kV Common 1kV Differential Surge 1KV CM Power Frequency Magnetic Field 30 A m 70 V for 0 5 period Voltage Dips 40 V for 5 50 periods 70 V for 25 periods Voltage I...

Страница 11: ...gh the bracket Slide the bracket toward the front of the case and remove Insert meter into the panel cutout Install mounting brackets and tighten the screws against the panel To achieve a proper seal...

Страница 12: ...at www predig com meterviewpro MeterView Pro Installation 1 Connect one end of the provided USB cable to the meter and the other end to the computer The computer will automatically install the driver...

Страница 13: ...ent and the meter Power Connections Power connections are made to a two terminal connector labeled POWER The meter will operate regardless of DC polarity connection The and symbols are only a suggeste...

Страница 14: ...ion provides sufficient lead wire compensation to obtain accurate readings even with long leads Figure 11 Thermocouple Input Connections Figure 12 Three Wire RTD Input Connections Figure 13 Two Wire R...

Страница 15: ...DC Loads Protection RC Networks Snubbers Available from Precision Digital RC networks are available from Precision Digital and should be applied to each relay contact switching an inductive load Part...

Страница 16: ...s the interlock relay feature reassigns one or more alarm control relays for use as interlock relay s Interlock contact s are wired to digital input s and activate the interlock relay This feature is...

Страница 17: ...eter Front Panel Buttons and Status LED Indicators Button Symbol Description LED Status Menu 1 8 Alarm 1 8 indicator Right arrow F1 1 8 M Flashing Relay in manual control mode Up arrow F2 M Manual con...

Страница 18: ...nual reset only after alarm condition has cleared Altern Alternate Set relay for alternation control Sanmpl Sampling Set relay for sampling operation Display Functions Messages Display Parameter Actio...

Страница 19: ...3 Set or enter Password 3 unloc Unlocked Program password to lock meter locd Locked Enter password to unlock meter 99999 9999 Flashing display Overrange condition Underrange condition Open Open Open s...

Страница 20: ...Up arrow button to scroll through choices Press the Menu button to exit at any time Setting the Input Signal Input Enter the Input menu to set up the meter to accept thermocouple tc or RTD rtd inputs...

Страница 21: ...to give the best performance under various lighting conditions Select intensity 8 for outdoor applications The default intensity is 6 Setting the Display Units or Custom Tags units Enter the display...

Страница 22: ...feature enabled 2 Set point 3 Reset point 4 Fail safe operation a On enabled b Off disabled 5 Time delay a On delay 0 999 9 seconds b Off delay 0 999 9 seconds 6 Relay action for loss break of input...

Страница 23: ...nd Off time delays may be programmed for each relay between 0 and 999 9 seconds The relays will transfer only after the condition has been maintained for the corresponding time delay The On time delay...

Страница 24: ...point the relay trips and the sample time starts After the sample time has elapsed the relay resets The cycle repeats every time the set point is crossed going up for high alarms and going down for lo...

Страница 25: ...gh or low temperature alarms control applications such as simple on off temperature control There are four basic ways the relays can be used 1 High or Low Alarms with Latching or Non Latching Relays 2...

Страница 26: ...n Relay terminology for following tables Terminology Relay Condition On Alarm Tripped Off Normal Reset Ack Acknowledged The On and Off terminology does not refer to the status of the relay s coil whic...

Страница 27: ...the ACK button or the assigned digital input is closed all relays programmed for manual reset are acknowledged Setting Up the Interlock Relay Force On Feature Relays 1 4 can be set up as interlock re...

Страница 28: ...ed with a decimal point regardless of the decimal point selection for the temperature display For instructions on how to program numeric values see Setting Numeric Values page 19 Reset Menu reset The...

Страница 29: ...ress the Enter button while the message is being displayed and enter the correct password to gain access to the menu Note After exiting the programming mode the meter returns to its password protected...

Страница 30: ...Action Setting Parity Parity Select parity Even Odd or None with 1 or 2 stop bits t byt Time byte Set byte to byte timeout SelEct Select Enter the Select menu AoutPr Analog output programmin g Program...

Страница 31: ...of accuracy to the calibration The scale can be changed to the Celsius scale after calibration is completed The meter will display temperature accurately in any scale The following table shows the re...

Страница 32: ...ion Digital s expansion modules cables or meters to the RJ45 M LINK connector Otherwise damage will occur to the equipment and the meter Note More detailed instructions are provided with each optional...

Страница 33: ...n be used to trigger external alarms or lights to indicate these specific events Function Keys Digital I O Available Settings Refer to the following table for descriptions of each available function k...

Страница 34: ...reference enter the room temperature e g 25 0 and press Enter 7 The meter displays 0mv apply an input equal to 0 000 mV and press Enter The display flashes while accepting the 0 mV input press Enter...

Страница 35: ...default settings for F1 F2 and F3 F4 Operation A digital input F4 is standard on the meter This digital input is programmed identically to function keys F1 F2 and F3 The input is triggered with a con...

Страница 36: ...see Advanced Features Menu Display Messages on page 30 Determining Software Version To determine the software version of a meter 1 Go to the Diagnostics menu diAG and press Enter button 2 Press Up ar...

Страница 37: ...relay 2 Off 2 0 0 sec On delay relay 3 On 3 0 0 sec Off delay relay 3 Off 3 0 0 sec On delay relay 4 On 4 0 0 sec Parameter Display Default Setting Off delay relay 4 Off 4 0 0 sec Input break relay 1...

Страница 38: ...onse is too slow Check filter and bypass values Display reading is not accurate Check 1 TC wire type 2 Input selection in Setup menu Display does not respond to input changes reading a fixed number Ch...

Страница 39: ...le technical standards and when applicable or required a European Union notified body certification Standards EN 55022 2003 EN 61000 6 2 2001 EN 61010 1 2001 EN 61326 2006 The standards EN 55022 2003...

Страница 40: ...predig com LIM7000_I SFT063 Ver 4 000 up 05 20 Contact Precision Digital Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com Sales Support Call 800 343 1001 o...

Отзывы: