P
RO
V
U
™ PD7000 Temperature Meter
Instruction Manual
30
Advanced Features Menu
To simplify the setup process, functions not needed
for most applications are located in the
Advanced
Features
menu.
Press and hold the Menu button for three seconds to
access the advanced features of the meter.
Advanced Features Menu &
Display Messages
The following table shows the functions and
messages of the
Advanced Features
menu in the
order they appear in the menu.
Display
Parameter
Action/Setting
Adjust
Adjust
Set adjust value to
offset temperature
reading
T cal
T Cal
Enter temperature
calibration menu
T lo
T Low
Apply the low
temperature input
(e.g. 32
°
F)
T Hi
T High
Apply the high
temperature input
(e.g. 1000
°
F)
rtdtot
RTD Total
Set the number of
RTDs connected in
parallel to obtain
average temperature
filtEr
Filter
Set noise filter value
bypASs
Bypass
Set filter bypass value
Round
Round
Set the rounding value
for display variables
seriAl
Serial
Set serial
communication
parameters
SlaV Id
Slave ID
Set slave ID or meter
address
baud
Baud rate
Select baud rate
Tr dLY
Transmit
delay
Set transmit delay for
serial communication
Display
Parameter
Action/Setting
Parity
Parity
Select parity
Even, Odd, or None
with 1 or 2 stop bits
t-byt
Time byte
Set byte-to-byte
timeout
SelEct
Select
Enter the Select menu
AoutPr
Analog
output
programmin
g
Program analog output
parameters
Source
Source
Select source for the
4-20 mA output
O-rang
Overrange
Program mA output for
display overrange
u-rang
Underrange
Program mA output for
display underrange
Break
Loop Break
Set relay condition if
loop break detected
nmAx
Maximum
Program maximum mA
output allowed
nmin
Minimum
Program minimum mA
output allowed
Calib
Calibrate
Calibrate 4-20 mA
output (internal
reference source used
for scaling the output)
4 nmA
4 mA output
Enter mA output value
read by milliamp meter
with at least 0.001 mA
resolution
20 nmA
20 mA
output
Enter mA output value
read by milliamp meter
with at least 0.001 mA
resolution
user
User I/O
Assign function keys
and digital I/O
F1
F1 function
key
Assign F1 function key
F2
F2 function
key
Assign F2 function key
F3
F3 function
key
Assign F3 function key
F4
F4 function
Assign F4 function
(digital input)
dI I
Digital input
1
Assign digital input
1- 8, if expansion
modules are
connected
dO 1
Digital output
1
Assign digital output
1- 8, if expansion
modules are
connected
ICal
Internal temp
calibration
Enter internal
temperature calibration
(used for factory
calibration only)